AI scribe notes should be evaluated for safety by having clinicians compare each note against the source audio, or a verified transcript, checking in both directions for unsupported content and for omissions, grading every error by how much it could change diagnosis or management, and reporting safety-critical errors separately rather than letting a good overall score average them away.
At DeepLabel, specialist clinicians from the BiteLabs network of 30,000+ UK and US clinicians work with PhD data scientists to agree the rubric, run the clinician review and produce the error and agreement statistics, in your own tools, or in ours. We verify each clinician's registration and experience, and match reviewers to the consultation setting. Need an independent evaluation of an AI scribe before deployment or procurement? Book a scoping call to talk through your product.
What makes an AI scribe note unsafe?
An AI scribe note is unsafe when it contains an error that could change diagnosis or management and the signing clinician does not catch it before relying on it. The errors that matter most are often the ones a reviewer is least likely to notice: omissions, and single-word changes such as a flipped negation inside an otherwise fluent note.
An ambient scribe does not harm a patient by itself. The harm comes when a clinician signs a note, relies on it, and the note is wrong in a way they did not catch. So what you are really evaluating is not the raw accuracy of a transcription model, but the risk that an error survives the clinician's review and reaches the record.
Two features of note review make certain errors more dangerous than others. Additions are errors of recognition: a fabricated finding or a wrong dose is on the page, and a careful reviewer can see it. Omissions are errors of recall: to notice that the safety-netting advice or the family history is missing, the clinician has to remember it was said, often hours later across a full clinic. That is a much harder task, which is why omissions tend to be common and easy to miss. In a simulated study of two commercial scribes, Biro and colleagues found 127 errors across 31 of 44 notes, a mean of 2.9 per note, with omissions the most frequent type; the authors suggest they may also be the hardest for clinicians to spot, because finding them depends on recall.
The second feature is that a note can be fluent, well-structured and almost entirely correct, and still contain one statement that changes care. A single flipped negation, "no chest pain" recorded as "chest pain", can redirect a whole consultation. So safety evaluation cannot be a single average: it has to find the rare, serious error inside a pile of good notes.
An AI scribe evaluation framework: the recommended approach
A practical AI scribe evaluation framework moves from defining intended use through source-grounded reference data, a severity rubric, representative sampling, calibrated clinician review, measurement, separate safety-critical reporting and ongoing monitoring. This is DeepLabel's recommended approach, not a regulatory standard and not a method prescribed by NHS England or the MHRA; treat it as a structure to adapt to your product's risk.
| Step | What it involves and the output |
|---|---|
| 1. Define intended use and clinical risks | Agree what the scribe is for, its care settings and the hazards that matter most. Output: Intended-use statement and a shortlist of key clinical hazards. |
| 2. Establish source-grounded reference data (audio or verified transcript) | Fix the reference each note is judged against: the audio, or a transcript verified against it. Output: A trustworthy source of truth for every sampled consultation. |
| 3. Define an error taxonomy and severity rubric | Write the error types, severity definitions and a safety-critical checklist, with worked examples. Output: A written rubric and annotation guidelines reviewers can apply consistently. |
| 4. Sample representative and challenging consultations | Draw a representative sample across settings and speakers, plus a separate enriched set of hard cases. Output: A sampling plan with representative and enriched strata reported apart. |
| 5. Conduct calibrated clinician review | Clinicians matched to setting grade each note against the source in both directions, after calibration. Output: Graded notes with errors typed and severity-rated. |
| 6. Measure errors and inter-rater reliability | Compute error rates by unit, with confidence intervals, and agreement by error type. Output: Error statistics and reliability figures with their uncertainty. |
| 7. Report safety-critical failure rates separately | Give safety-critical errors their own rate and examples, never averaged into a pass rate. Output: A standalone safety-critical section with examples and limitations. |
| 8. Run regression testing and ongoing monitoring | Re-test every product change against a fixed set, and monitor live use after deployment. Output: A version-controlled regression set and a monitoring plan. |
Grading against the source in step 2 is not optional. How far steps 4 to 6 go, including how much double review, calibration and adjudication, is agreed at scoping and scaled to the product's risk: a higher-risk scribe may warrant a larger double-reviewed subset and formal adjudication, where a lower-risk tool may need less.
What errors should an AI scribe evaluation look for?
An AI scribe evaluation should look for omissions, hallucinated or fabricated content, negation errors, misattribution, wrong laterality, dose, units, timings or values, contextual errors, and coding or summary errors, with safety-critical items such as allergies, medications and red flags checked on every note.
To grade notes consistently, agree a written taxonomy of what can go wrong before anyone starts reviewing. A workable one for scribe notes separates error types so that a cosmetic slip and a dangerous miss are never counted the same:
- Omissions: clinically relevant content that was in the consultation but is missing from the note, such as a dropped safety-netting instruction or a follow-up plan.
- Hallucinated or fabricated content: statements in the note that the consultation does not support, including findings or symptoms that were never discussed.
- Negation errors: a positive turned negative or the reverse, such as "no chest pain" becoming "chest pain". Asgari and colleagues found negation errors were about 30% of hallucinations in their study and called them the most concerning subtype.
- Misattribution: the wrong speaker or source, such as a relative's history recorded as the patient's own, or advice the clinician gave attributed to the patient.
- Wrong laterality, dose, units, timings or values: a left knee written as the right, 5 mg written as 50 mg, or a blood pressure transposed.
- Contextual errors: a hypothetical stated as fact (a patient's "I'm scared it might be cancer" recorded as a diagnosis), uncertainty turned into certainty, or a causal link invented between two things that were only mentioned together.
- Coding and summary errors: clinical codes not supported by what was said, or problem-list and summary entries that distort the consultation.
Over the top of that, name the safety-critical items explicitly and check them on every note: allergies, medications and doses, red-flag symptoms, safety-netting advice, follow-up and referrals, and any safeguarding concern. These are the items where a single error is most likely to cause harm, so they are tracked and reported on their own, not folded into a general error count.
Note quality is worth measuring too, but keep it separate from safety. Structure, usability and concision affect whether the next clinician can use the note; that is not the same question as whether the note is true. The Physician Documentation Quality Instrument (PDQI-9), a validated nine-item scale for note quality (up to date, accurate, thorough, useful, organised, comprehensible, succinct, synthesised, consistent), is a reasonable instrument to borrow for the quality half, as long as a good quality score does not stand in for a safety judgement.
How should AI scribe errors be graded for severity?
AI scribe errors should be graded by clinical impact, not just counted. The simplest workable cut is the one Asgari and colleagues used: an error is major if it could change diagnosis or management if left uncorrected, and minor otherwise. That single distinction does most of the work, because it separates the wording slip from the error that moves care.
Grading individual errors as major or minor during an evaluation is not the same as the formal clinical risk management required under the NHS standards DCB0129 (for manufacturers of health IT systems) and DCB0160 (for the organisations deploying and using them). Per NHS England's ambient scribing guidance, DCB0160 documentation includes a safety case, a hazard log and a monitoring framework, and adopters should ask for the supplier's DCB0129 documentation, the clinical safety case report; hazards are assessed by a clinical safety officer. Evaluation findings are evidence that can inform the hazard log and the safety case, but they do not replace them. Asgari and colleagues based their safety assessment on the framework used for evaluating a medical device, estimating the likelihood of an error together with the severity of harm it could cause; that is a useful way to prioritise errors, not a substitute for the manufacturer's risk management.
One rule matters more than the rest: a single safety-critical error must never be averaged away by a good overall score. A product can post a low mean error rate and still miss one allergy in a hundred notes, and that one note is the one that matters. Report safety-critical errors separately, with their own rate and their own examples, and resist any summary that lets a high pass rate hide them. Severity grading also shows where errors cluster: Asgari and colleagues found major hallucinations most common in the Plan section, exactly where a wrong statement is most likely to drive an action.
How do you review AI scribe notes against the source?
You review AI scribe notes by having a clinician compare each note against the source, statement by statement, in both directions: note to source to catch unsupported content, and source to note to catch omissions. Both passes are needed.
Going from the note to the source catches unsupported content: for each statement in the note, is it actually supported by what was said? Going from the source to the note catches omissions: for each clinically relevant thing in the consultation, did it make it into the note? You need both passes.
The source has to be the audio, or a transcript that has itself been verified against the audio. Grading against the clinician's memory does not work, because recall is exactly the faculty that fails on omissions, and an unverified transcript simply moves the error one step upstream: if the transcript is wrong, a note that faithfully follows it will look correct. This is the principle behind our AI scribe evaluation work and our guide to evaluating a clinical LLM: define the reference, then measure against it, not against an impression.
Give reviewers a written rubric and annotation guidelines, with the taxonomy, the severity definitions and worked examples of each. Then calibrate on a pilot set before the main review: a small, varied batch that several clinicians grade, with the disagreements read back to tighten the definitions. Ambiguous guidelines are the most common reason reviewers diverge, and a pilot is where you find and fix them early, before you have committed the main volume to an inconsistent standard.
How should you measure AI scribe error rates?
There is no single AI scribe error rate: the number depends on the unit of analysis, on whether you measure the draft note or the signed note, and on how you report uncertainty. Report rates per sentence, per note and per clinically relevant item, state whether they are on draft or signed notes, and give a confidence interval for every rate.
The unit of analysis changes the number, so it has to be stated. Asgari and colleagues report rates per sentence: 191 of 12,999 note sentences contained a hallucination (1.47%), across 450 notes, and omissions were 1,712 of 49,590 transcript sentences (3.45%). Biro and colleagues report per note: 127 errors across 44 draft notes, a mean of 2.9 errors per note, with 31 of 44 notes (70%) containing at least one error. The two cannot be compared directly: as a labelled illustration, 191 hallucinated sentences spread across 450 notes is roughly 0.4 hallucinated sentences per note, which is not the same quantity as an all-errors-per-note count. The designs differ too: Asgari evaluated LLM notes from 450 transcript-note pairs built on PriMock mock primary care consultations across 18 configurations, while Biro tested 2 commercial products across 11 scripted simulated encounters, so the units are not the only thing that differs.
We recommend also reporting errors per clinically relevant item discussed, for example per medication, allergy, red flag or follow-up action, because that denominator sits closest to clinical risk; and reporting the proportion of notes with at least one major error. Present these as recommendations: they describe how we think rates are best reported, not something either study did.
Draft-note errors are not the same as errors that survive clinician review. Both studies measured draft output before any clinician correction: Biro explicitly evaluated draft notes, and Asgari evaluated LLM-generated notes against transcripts. Neither measured how many errors survive the signing clinician's review, and it is those surviving errors that actually reach the record. Measuring them needs a different design, such as comparing the signed note, after clinician editing, against the source. Both measures are useful; report them separately and label which is which.
Finding no serious errors does not establish safety. Zero major errors in a sample does not show a product is safe; it only bounds how common such errors are likely to be. The rule of three is a standard statistical rule of thumb here: if 0 events are seen in n independent cases, an approximate one-sided 95% upper confidence bound on the true rate is 3/n (Hanley and Lippman-Hand, JAMA 1983). As an illustration only, zero major errors in 300 reviewed notes is still compatible with a true rate of up to about 1 in 100 notes (3/300). It assumes independent, representative cases, so it says nothing about consultation types that were not sampled.
Report confidence intervals for every rate, so a reader can see the precision behind each number rather than a point estimate on its own.
How many consultations should you sample?
There is no single correct number of consultations to sample: the sample size follows from the error rates you need to estimate and the confidence you need in them, and it should be fixed in a plan before review starts. A rare safety-critical error needs far more notes to measure reliably than an overall error rate does, and the sample must also span the specialties, settings and speakers you serve.
What you sample determines what your results mean. A scribe that performs well on clean, single-speaker follow-ups in one specialty can fail on a noisy new patient with an interpreter, and an evaluation drawn only from the easy cases will not show it. Sample across the dimensions that actually move performance:
- specialty and care setting;
- consultation type: face-to-face, telephone and video, and new versus follow-up;
- speaker factors: accents, regional dialects, English as a second language, interpreters, speech impairments and multiple speakers;
- conditions: background noise and long or complex consultations.
NHS England's guidance asks adopters to probe the same ground: whether speakers are tagged in the transcript, how the product handles UK-specific terminology, and speaker characteristics such as age, accent, dialect and disorders like aphasia. It is explicit that these products may vary with some accents or regional dialects, and where speakers use English as a second language or are affected by speech disorders. Enrich a sample with hard cases where that is efficient, but report the enriched cases separately from a representative sample, so an inflated hard-case error rate is not read as the typical one.
How do you measure reviewer agreement?
Reviewer agreement is measured by having two clinicians grade the same notes independently and quantifying how often they agree, per error type, with a statistic that stays stable when errors are rare. It tells you how much of the result reflects the product and how much reflects the reviewer.
A single reviewer is one clinical opinion, and clinicians do not always agree on what counts as a clinically relevant omission or a major error. Where the stakes justify it, have a subset of notes reviewed independently by two clinicians, or review everything twice for the highest-risk material, and measure how often they agree. Report agreement per error type rather than as one figure: a pair might agree closely on fabricated content and poorly on what counts as a missing safety-net, and an average hides that.
Choose the statistic to fit the data. Cohen's kappa is the usual starting point for whether an error class is present, but it becomes unstable when an error type is rare and the ratings are heavily skewed, which is the norm for safety-critical errors. Gwet's AC1 is more stable under that skew, so quote it alongside kappa rather than relying on either alone, and have a senior clinician adjudicate disagreements so each note ends with one agreed judgement. None of this is standard on every project: calibration rounds, quality sampling as work comes in, independent double reads and senior adjudication are design choices made at scoping and scaled to the risk of the work.
How do you red-team an ambient scribe?
You red-team an ambient scribe by deliberately constructing or selecting the consultations most likely to break it, rather than waiting for a representative sample to surface rare failures. The targets come straight from the error taxonomy: dense negation, sound-alike drug names, laterality, overlapping speech, hypotheticals and utterances that push the product beyond its intended purpose.
A representative sample shows typical behaviour. Red-teaming goes looking for the failures a representative sample will rarely surface, by deliberately constructing or selecting the consultations most likely to break a scribe. The targets follow from the taxonomy:
- dense negation and double negatives;
- look-alike, sound-alike drug names, and numbers, doses and units spoken quickly;
- laterality, where left and right are both mentioned;
- overlapping speech and a third party in the room, such as a parent or carer;
- hypotheticals and a patient who corrects themselves part way through;
- utterances that read like instructions to the product, and requests that push it beyond its intended purpose, such as a clinician asking "what do you think the diagnosis is?".
That last category is one NHS England calls out directly: generative products may introduce functions unintentionally, a user might get an output beyond the product's intended purpose, and suppliers should have guardrails against users instructing the product to suggest diagnoses. The MHRA names hallucination as a well-known behaviour of generative models and expects manufacturers to mitigate risk by design first, or with other protective measures, before considering warnings. Red-teaming is how you find out whether those guardrails hold before a real consultation tests them, and it is the same discipline as our medical red-teaming work: probe the edges on purpose, so a patient is not the one who finds them.
How should AI scribes be monitored after deployment?
AI scribes should be monitored continuously after deployment, because the product changes: models are updated, prompts are tuned and templates are edited, and any of these can move the error profile. Combine regression testing against a fixed, version-controlled test set on every change with ongoing monitoring of live use, including audits of signed notes against the source and review of incident reports.
A one-off evaluation describes the product on the day you tested it. Safety is a property over time, and a changed product can behave very differently, sometimes sharply so. Asgari and colleagues found that prompt and workflow changes altered error rates substantially, and that one change intended to help, a fact-extraction step before the note was written, actually increased major errors. The lesson is blunt: every change needs re-evaluation, with regression testing against a fixed, version-controlled test set so you can see what moved.
Alongside regression testing, run ongoing monitoring of live use. NHS England asks deploying organisations to ensure output accuracy through ongoing audits of clinical documentation and reviews of incident reports and system performance, and is specific that the organisation's monitoring data should be collected and analysed independently of the manufacturer's. Useful signals include audits of signed notes against the source, incident reports routed through local systems and the Learn from Patient Safety Events service, device concerns reported to the MHRA through the Yellow Card scheme, and trends in clinician edit rates and error types. Watch subgroups for drift, and watch for automation bias: if clinicians edit less as they come to trust the tool, the final safety net quietly weakens. The NHS England and MHRA joint letter reinforces the contractual side, advising boards to secure rights to monitor performance and manage changes in functionality, including at AI model update milestones.
Who should review AI scribe notes?
AI scribe notes should be reviewed by clinicians working in the same specialty and care setting as the consultations, with current registration and relevant recent experience, and matched to country where drug names, units and care pathways differ. A note from a psychiatry clinic and one from an emergency department make different demands, and a reviewer outside the setting will miss context-specific errors.
Setting matters because what counts as a clinically relevant omission, or a plausible dose, differs between specialties. Match on country where it matters, because drug names, units, dosing conventions and care pathways differ between the UK and the US, and a reviewer applying the wrong convention will either miss real errors or flag correct notes as wrong. We verify each clinician's registration and experience and match them to the setting; reviewers are UK and US, with access set per project.
Review is a team effort with clear ownership. A clinical safety officer owns the hazard log and the safety case. Clinicians make the clinical judgements. PhD data scientists own the sampling, the statistics and the agreement analysis. An LLM-as-judge can help triage at scale, flagging likely omissions or unsupported statements for a clinician to confirm, but only once it has been calibrated against clinician grades, and it should never make the release decision on its own. We set out how to validate an autograder against clinicians in our guide on calibrating an LLM judge for clinical outputs; the short version is that a judge with a high overall agreement score can still miss the rare safety-critical errors, so you measure it per severity before you trust it.
How do you validate an ambient scribe before NHS deployment?
Validating an ambient scribe before NHS deployment means building the evidence NHS England's guidance asks for, not following a single prescribed protocol: NHS England does not mandate one evaluation method. In practice that evidence is a severity-graded clinician review against the source, the supplier's safety documentation, and a plan for ongoing monitoring, assembled to match the product's intended use and risk.
The method above does not replace the official guidance; it generates the evidence that guidance asks for. NHS England's guidance on AI-enabled ambient scribing products, aimed at CIOs and CCIOs in England, asks adopters to assign a clinical safety officer, identify key clinical hazards (it says these may include missing critical information, incorrect information or context, or delayed outputs), complete DCB0160 documentation and a data protection impact assessment, and confirm the supplier has completed DCB0129. It also asks for the supplier's evidence on the product's accuracy, including validation from prospective user groups such as clinicians on the rate of errors, omissions and other dysfunctions. A severity-graded clinician review against the source is one of the most direct ways to produce that evidence, whether the supplier commissions it or the adopter does.
The MHRA's guidance on ambient voice technology products, which applies in Great Britain, turns on intended purpose rather than the technology. Intended purpose is defined by the claims made in the instructions for use, the labelling and the promotional materials, including websites and adverts. A product that solely transcribes, summarises for clinician review before saving, structures a transcript, suggests codes from explicitly mentioned terms, or drafts a letter for review is, on the MHRA's examples, not a medical device. The moment the intended purpose extends to diagnosis or treatment, or the product acts without clinician review, it can become one. A general "not for diagnosis" disclaimer does not prevent device status if medical claims are made or implied elsewhere, including in marketing. The guidance applies to products placed on the Great Britain market under the UK Medical Devices Regulations 2002; the MHRA notes that EU MDR applies in Northern Ireland under the Windsor Framework.
| Intended purpose (MHRA examples) | Device status in the MHRA example |
|---|---|
| Transcribes a consultation for clinician review | Not a medical device |
| Summarises a consultation for clinician review before it is saved to the record | Not a medical device |
| Formats a transcript or summary into structured data for clinician confirmation | Not a medical device |
| Suggests clinical codes from explicitly mentioned terms, for review | Not a medical device |
| Drafts a letter or discharge summary for clinician review | Not a medical device |
| Offers "generated insights" suggesting diagnoses or treatment, despite a disclaimer | A medical device |
| Summarises only, but is marketed as guiding diagnosis and treatment | A medical device |
| Saves notes and orders follow-up tests without clinician review | A medical device |
| Drafts a report and suggests diagnoses for clinician review | A medical device |
These rows are the MHRA's own worked illustrations; a specific product's status must be assessed against its own intended purpose, and the MHRA is explicit that the guidance does not change the law.
If a product is a device, it is classified under the UK Medical Devices Regulations 2002, and the MHRA guidance discusses the active-device rules, with class IIa and above needing certification from an approved or notified body. Keep that at a high level: the classification of a specific product sits with its manufacturer and their regulatory advisers. The MHRA, NHS England and their joint letter are consistent on the point that matters most for safety, which the MHRA states plainly: clinicians remain responsible for reviewing and verifying AI-generated transcripts, summaries and other outputs before they are used in patient care. Manufacturers should also reassess with every product change, because an added feature can alter the intended purpose and with it the regulatory status.
What evidence should an AI scribe supplier provide?
An AI scribe supplier should be able to provide, at minimum, their DCB0129 clinical safety case report, evidence on the product's accuracy from prospective user groups, a clear statement of intended use and usage policy with guardrails, and confirmation that the product has been correctly considered under medical device regulation. These are the points NHS England's guidance directs adopters to check.
The points in NHS England's guidance are:
- the supplier's DCB0129 documentation, that is, the clinical safety case report;
- evidence on the accuracy of the product, including validation from prospective user groups such as clinicians, on the rate of errors, omissions and other dysfunctions;
- the product's intended use and usage policy, with guardrails against users instructing the product to suggest diagnoses or identify missing consultation components, especially where it is not a registered medical device;
- that the product has been correctly considered under medical device regulation, and complies with it where it applies; whether it is a device depends on its intended purpose, as set out in the MHRA's guidance.
Alongside the guidance points, we suggest asking a set of questions that get behind a headline accuracy figure. These are our recommendations, not a regulatory requirement:
- what unit the error rates use: per sentence, per note or per clinically relevant item;
- whether the rates are on draft notes or on signed notes after clinician editing;
- how the sample was drawn, and across which specialties, accents and consultation types;
- who reviewed the notes, and with what measured agreement;
- how safety-critical errors were reported, and whether separately from the overall rate;
- whether any part of the evaluation was carried out independently of the developer.
NHS England's guidance also points adopters to its AVT Supplier Registry, a national self-certified registry of ambient scribing suppliers with a repository of supporting evidence that NHS organisations can use in local procurement and assurance. And it asks deploying organisations to collect and analyse their own monitoring data independently of the manufacturer's, so a supplier's own figures are a starting point, not the whole picture.
Why use an independent AI scribe evaluation?
An independent AI scribe evaluation gives you a clinician-led assessment of your notes that is separate from the developer's own internal testing, which matters because NHS England asks deploying organisations to collect and analyse monitoring data independently of the manufacturer. NHS England does not, by itself, require an independent pre-deployment evaluation, but independent evidence can carry more weight, and is easier to defend, when a safety case is scrutinised.
This is the work DeepLabel does. Specialist clinicians from the BiteLabs network of 30,000+ UK and US clinicians are matched to your specialty, care setting and country, with reviewers in the UK and US and access set per project, and each clinician's registration and experience verified. The annotation rubric and guidelines are clinician-developed and agreed with you; calibration and adjudication are design options we scope with you rather than fixed steps on every project. Clinical data quality assurance is built into the review, and PhD data scientists run the quantitative analysis of error types, error rates and reviewer agreement. We work in your own tools, or in ours.
The same approach runs across our wider services: clinical AI evaluation for models and outputs beyond scribe notes, medical LLM evaluation where the underlying model is the focus, and medical data annotation for the reference data an evaluation rests on.
One boundary is worth stating clearly. An evaluation like this provides clinical insight to support product development and to feed a safety case; it does not make a scribe safe, certify its safety or confer regulatory approval, and it does not replace formal validation. It is one part of a wider body of evidence, and a strong part, but not the whole.
Example: how an AI scribe safety evaluation could be structured
The following is an illustrative design showing how an AI scribe safety evaluation could be structured. It is not a real project, describes no client and contains no results: it only shows how the pieces above fit together.
- Scope: a scribe used in UK general practice and urgent care, with the intended use of summarising a consultation for clinician review before saving.
- Sample design: stratified by care setting and by consultation type, including face-to-face, telephone and video, and by speaker factors such as a range of accents, English as a second language and interpreter-mediated consultations, with a separately reported enriched set of hard cases.
- Reference: the consultation audio, or transcripts verified against the audio, as the source each note is judged against.
- Rubric: the error taxonomy, a major or minor severity grade, and a safety-critical checklist covering allergies, medications, doses, red flags, safety-netting and follow-up.
- Review: clinicians matched to the care setting, with a double-reviewed subset and adjudication if those are agreed at scoping.
- Metrics: error rates per sentence, per note and per clinically relevant item, safety-critical rates with confidence intervals, and reviewer agreement reported per error type.
- Reporting: a separate safety-critical section, subgroup results by setting and speaker factor, a worked example of each major error type, stated limitations, and a regression set retained for future versions.
If you are building an ambient scribe and want its notes evaluated for safety against the source, by clinicians matched to your settings and with the statistics to back the findings, that is what we do. Our AI scribe evaluation service grades every error by type and severity, and our broader clinical AI evaluation service extends the same approach across other outputs, models and versions, in your own tools, or in ours. Book a scoping call to talk it through.